PDF EndNote BibTex Kaynak Göster

Analytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detector

Yıl 2015, Cilt 2, Sayı 2, 11 - 22, 20.06.2015
https://doi.org/10.1501/nuclear_0000000008

Öz

In this paper, a direct analytical method is presented for calculating the absolute efficiency of an elliptical cylindrical detector in the case of an arbitrarily positioned (above the major axis a) point source. The absolute efficiency is required to determine the activity of an unknown radioactive source, taking into account the
attenuation of the gamma-ray photons. The validity of the derived analytical expressions was successfully confirmed by the comparison with some published data.

Kaynakça

  • Abbas, M., “A direct mathematical method to calculate the efficiencies of a parallelepiped detector for an arbitrarily positioned point source”, J. Radiat. Phys. Chem.60, 3 (2001a).
  • Abbas, M., “HPGe detector photopeak efficiency calculation including self-absorption and coincidence corrections for Marinilli beaker sources using compact analytical expressions”, J. Radiat. Phys. Chem. 54, 761 (2001b).
  • Abbas, M., Selim, Y., “Calculation of relative full-energy peak efficiencies of well-type detectors”, Nucl. Instrum. Methods Phys. Res. A480, 651 (2002).
  • Abbas, M., Selim, Y., Nafee, S., “A simple mathematical method to determine the efficiencies of log-conical detectors”, J. Radiat. Phys. Chem. 75 (7), 729-736 (2006).
  • Jiang, S., Liang, J., Chou, J., Lin, U., Yeh, W., "A hybrid method for calculating absolute peak efficiency of germanium detectors", Nucl. Instrum. Methods Phys. Res. A413, 281-292 (1998).
  • Lippert, J., “Detector-efficiency calculation based on point-source measurement”, Int. J. Appl. Radiat. Isot. 34 (8), 1097-1103 (1983).
  • Moens, L., Hoste, J., “Calculation of the peak efficiency of high-purity germanium detectors”, Int. J. Appl. Radiat. Isot. 34(8), 1085–105 (1983).
  • Nakamura, T., Suzuki, T., “Monte Carlo calculation of peak efficiencies of Ge(Li) and pure Ge detectors to voluminal sources and comparison with environmental radioactivity measurement”, Nucl. Instrum. Methods Phys. Res. 205 (1), 211-218 (1983).
  • Rieppo, R., “Calculated absolute full-energy peak efficiencies for true coaxial Ge(Li) detector in the photon energy region 0.1-0.3 MeV with different source-to-detector geometries”, Int. J. Appl. Radiat. Isot. 36 (11): 861-865 (1985).
  • Selim, Y, Abbas, M., “Source-detector geometrical efficiency”, J. Radiat. Phys. Chem. 44 (1), 1-4 (1994).
  • Selim, Y., Abbas, M., “Analytical calculations of gamma scintillators efficiencies. II: total efficiency for wide co-axial disk sources”, J. Radiat. Phys. Chem. 53, 15–19 (2000).
  • Selim, Y., Abbas, M., Fawzy, M., “Analytical calculation of the efficiencies of gamma scintillators. Part I: Total efficiency for coaxial disk sources”, J. Radiat. Phys. Chem. 53 (6), 589- 592 (1998).
  • Wang, T., Mar, W., Ying, T., Liao, C., Tseng, C., “HPGe detector absolute-peak-efficiency calibration by using the ESOLAN program”, Int. J. Appl. Radiat. Isot. 46 (9), 933–944 (1995).
  • Wang, T., Mar, W., Ying, T., Tseng, C., Liao, C., Wang, M., “HPGe detector efficiency calibration for extended cylinder and marinellibeaker sources using the ESOLAN program”, Int. J. Appl. Radiat. Isot. 48(1), 83–95 (1997).
  • Yalcin, S, Gurler, O, Kaynak, G, Gundogdu, O., “Calculation of total counting efficiency of a NaI(Tl) detector by hybrid Monte-Carlo method for point and disk sources”, Int. J. Appl. Radiat. Isot. 65, 1179–1186 (2007).

Yıl 2015, Cilt 2, Sayı 2, 11 - 22, 20.06.2015
https://doi.org/10.1501/nuclear_0000000008

Öz

Kaynakça

  • Abbas, M., “A direct mathematical method to calculate the efficiencies of a parallelepiped detector for an arbitrarily positioned point source”, J. Radiat. Phys. Chem.60, 3 (2001a).
  • Abbas, M., “HPGe detector photopeak efficiency calculation including self-absorption and coincidence corrections for Marinilli beaker sources using compact analytical expressions”, J. Radiat. Phys. Chem. 54, 761 (2001b).
  • Abbas, M., Selim, Y., “Calculation of relative full-energy peak efficiencies of well-type detectors”, Nucl. Instrum. Methods Phys. Res. A480, 651 (2002).
  • Abbas, M., Selim, Y., Nafee, S., “A simple mathematical method to determine the efficiencies of log-conical detectors”, J. Radiat. Phys. Chem. 75 (7), 729-736 (2006).
  • Jiang, S., Liang, J., Chou, J., Lin, U., Yeh, W., "A hybrid method for calculating absolute peak efficiency of germanium detectors", Nucl. Instrum. Methods Phys. Res. A413, 281-292 (1998).
  • Lippert, J., “Detector-efficiency calculation based on point-source measurement”, Int. J. Appl. Radiat. Isot. 34 (8), 1097-1103 (1983).
  • Moens, L., Hoste, J., “Calculation of the peak efficiency of high-purity germanium detectors”, Int. J. Appl. Radiat. Isot. 34(8), 1085–105 (1983).
  • Nakamura, T., Suzuki, T., “Monte Carlo calculation of peak efficiencies of Ge(Li) and pure Ge detectors to voluminal sources and comparison with environmental radioactivity measurement”, Nucl. Instrum. Methods Phys. Res. 205 (1), 211-218 (1983).
  • Rieppo, R., “Calculated absolute full-energy peak efficiencies for true coaxial Ge(Li) detector in the photon energy region 0.1-0.3 MeV with different source-to-detector geometries”, Int. J. Appl. Radiat. Isot. 36 (11): 861-865 (1985).
  • Selim, Y, Abbas, M., “Source-detector geometrical efficiency”, J. Radiat. Phys. Chem. 44 (1), 1-4 (1994).
  • Selim, Y., Abbas, M., “Analytical calculations of gamma scintillators efficiencies. II: total efficiency for wide co-axial disk sources”, J. Radiat. Phys. Chem. 53, 15–19 (2000).
  • Selim, Y., Abbas, M., Fawzy, M., “Analytical calculation of the efficiencies of gamma scintillators. Part I: Total efficiency for coaxial disk sources”, J. Radiat. Phys. Chem. 53 (6), 589- 592 (1998).
  • Wang, T., Mar, W., Ying, T., Liao, C., Tseng, C., “HPGe detector absolute-peak-efficiency calibration by using the ESOLAN program”, Int. J. Appl. Radiat. Isot. 46 (9), 933–944 (1995).
  • Wang, T., Mar, W., Ying, T., Tseng, C., Liao, C., Wang, M., “HPGe detector efficiency calibration for extended cylinder and marinellibeaker sources using the ESOLAN program”, Int. J. Appl. Radiat. Isot. 48(1), 83–95 (1997).
  • Yalcin, S, Gurler, O, Kaynak, G, Gundogdu, O., “Calculation of total counting efficiency of a NaI(Tl) detector by hybrid Monte-Carlo method for point and disk sources”, Int. J. Appl. Radiat. Isot. 65, 1179–1186 (2007).

Ayrıntılar

Birincil Dil İngilizce
Bölüm Articles
Yazarlar

M. ABBAS Bu kişi benim


S. HAMMOUD>


T. IBRAHİM Bu kişi benim


M. SAKR Bu kişi benim

Yayımlanma Tarihi 20 Haziran 2015
Yayınlandığı Sayı Yıl 2015, Cilt 2, Sayı 2

Kaynak Göster

Bibtex @ { ankujns22818, journal = {Journal of Nuclear Sciences}, issn = {2147-7736}, eissn = {2148-3981}, address = {}, publisher = {Ankara Üniversitesi}, year = {2015}, volume = {2}, number = {2}, pages = {11 - 22}, doi = {10.1501/nuclear\_0000000008}, title = {Analytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detector}, key = {cite}, author = {Abbas, M. and Hammoud, S. and Ibrahim, T. and Sakr, M.} }
APA Abbas, M. , Hammoud, S. , Ibrahim, T. & Sakr, M. (2015). Analytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detector . Journal of Nuclear Sciences , 2 (2) , 11-22 . DOI: 10.1501/nuclear_0000000008
MLA Abbas, M. , Hammoud, S. , Ibrahim, T. , Sakr, M. "Analytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detector" . Journal of Nuclear Sciences 2 (2015 ): 11-22 <http://jns.ankara.edu.tr/tr/pub/issue/1884/22818>
Chicago Abbas, M. , Hammoud, S. , Ibrahim, T. , Sakr, M. "Analytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detector". Journal of Nuclear Sciences 2 (2015 ): 11-22
RIS TY - JOUR T1 - Analytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detector AU - M. Abbas , S. Hammoud , T. Ibrahim , M. Sakr Y1 - 2015 PY - 2015 N1 - doi: 10.1501/nuclear_0000000008 DO - 10.1501/nuclear_0000000008 T2 - Journal of Nuclear Sciences JF - Journal JO - JOR SP - 11 EP - 22 VL - 2 IS - 2 SN - 2147-7736-2148-3981 M3 - doi: 10.1501/nuclear_0000000008 UR - https://doi.org/10.1501/nuclear_0000000008 Y2 - 2022 ER -
EndNote %0 Journal of Nuclear Sciences Analytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detector %A M. Abbas , S. Hammoud , T. Ibrahim , M. Sakr %T Analytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detector %D 2015 %J Journal of Nuclear Sciences %P 2147-7736-2148-3981 %V 2 %N 2 %R doi: 10.1501/nuclear_0000000008 %U 10.1501/nuclear_0000000008
ISNAD Abbas, M. , Hammoud, S. , Ibrahim, T. , Sakr, M. . "Analytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detector". Journal of Nuclear Sciences 2 / 2 (Haziran 2015): 11-22 . https://doi.org/10.1501/nuclear_0000000008
AMA Abbas M. , Hammoud S. , Ibrahim T. , Sakr M. Analytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detector. Journal of Nuclear Sciences. 2015; 2(2): 11-22.
Vancouver Abbas M. , Hammoud S. , Ibrahim T. , Sakr M. Analytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detector. Journal of Nuclear Sciences. 2015; 2(2): 11-22.
IEEE M. Abbas , S. Hammoud , T. Ibrahim ve M. Sakr , "Analytical Formulae to Calculate the Total Efficiency of an Arbitrarily Positioned Point Source by an Elliptical Cylindrical Detector", Journal of Nuclear Sciences, c. 2, sayı. 2, ss. 11-22, Haz. 2015, doi:10.1501/nuclear_0000000008